Hand-held Analyzer Type 2250
Instruction Manual
48
4.12.1 Environmental Stabilisation Time
Typical stabilisation time after change in environmental conditions is 1 minute.
When the instruments are moved from a warm environment with high humidity, to a colder environ-
ment, care should be taken not to produce condensation inside the instruments. In this case, much
longer stabilisation periods may be necessary.
4.12.2 Temperature
Operating temperature range: –10 to +50ºC (+14 to +122ºF)
Storage temperature range: –25 to +70ºC (–13 to +158ºF)
4.12.3 Humidity
Operating humidity range: 0% < RH < 90%, providing there is no condensation.
4.12.4 Vibration
Vibration sensitivity (< 1000Hz), measured with Z-weighting: Typically 62.5 dB for 1 ms
–2
, max
65 dB for 1 ms
–2
.
4.12.5 Magnetic Fields
Max. sensitivity to power line (50/60 Hz) magnetic field strength of 80 A/m:
4.12.6 Immunity to Power and Radio-frequency Fields
Conforms to IEC 60651, IEC 60804 and IEC 61672–1 down to 50 dB.
Fig.4.31
The most sensitive
direction
Table 4.13
Magnetic Fields
Configuration Most Sensitive Direction
Max. Sensitivity
A-weighted
(dB)
C-weighted
(dB)
Z-weighted
(dB)
Type 2250 alone
Not
detectable
<19 <20
Type 4189 and
Type ZC 0032
alone
Not
detectable
<26 <27
040141
H - field
E - field
Direction of
magnetic field
040140
BE171211.book Page 48 Thursday, May 27, 2004 1:01 PM