B&K 2250 Light Manuel d'utilisateur Page 25

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 122
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 24
Hand-held Analyzer Type 2250
Instruction Manual
20
4.4 Reference Conditions for Acoustic Calibration
Reference range: Only one range exists and this is the reference range
Reference sound pressure level: 94.00 dB re 20 µPa
Reference frequency: 1kHz
4.5 Microphone
Microphone Type 4189 and Preamplifier Type ZC 0032:
Type: Prepolarized Free-field ½ Condenser Microphone
Nominal Open Circuit Sensitivity: 50 mV/Pa, (corresponding to –26 dB re 1 V/Pa) ± 1.5 dB
Capacitance: 13 pF (at 250 Hz)
Nominal preamplifier attenuation: 0.25 dB
Extension cables between Type ZC0032 and Type 2250: Up to 100 m without affecting measure-
ments. Note: EMC is only tested with a 10 m cable (AO 0442)
Microphone reference point: The centre of the front surface of the microphone protection grid. With
UA 1404 mounted, this point can be difficult to find. Therefore, it is marked on the windscreen
Reference direction of sound incidence: See Fig.4.14 to Fig.4.15
4.6 Frequency Response
The frequency responses are given in tabular form in Appendix A and in graphical form in this section.
The Expanded Uncertainties of Measurement in the tables are the two-sigma limits maintained in the
Brüel & Kjær production for the specific product. This means that the correct value lies in the range of
the measured value plus/minus the Expanded Uncertainties of Measurement with a probability of 95%.
Note: This range is not related to the production spread. The range is for a specific typical unit. The
factory acceptance tests at Brüel & Kjær ensure that the ranges for all units are within the IEC 61672–1
limits.
The IEC 61672–1 limits, reduced by the Maximum Expanded Uncertainties of Measurement from Ap-
pendix A of IEC 61672–1, are also drawn in the graphs. These Maximum Expanded Uncertainties of
Measurement are the maximum uncertainties that a test organisation may have on its measurements
when it performs conformance tests according to IEC 61672.
4.6.1 Electrical Frequency Responses
The electrical frequency response depends on the selected frequency weighting, see Fig.4.1.
The Z-weighted electrical frequency response is given in the "Electrical Response" column of the
appropriate tables of Table A.2 to Table A.9 in Appendix A.
The electrical frequency responses are valid both for introduction of the electrical signal through the
recommended means to substitute the microphone with an electrical input facility (see section 3.4) and
for the rear ‘Input’ socket.
Z-weighted frequency range: ±0.15dB, 11Hz to 22kHz
Extended low frequency range: ±0.15dB, 4Hz to 22kHz
BE171211.book Page 20 Thursday, May 27, 2004 1:01 PM
Vue de la page 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 121 122

Commentaires sur ces manuels

Pas de commentaire