50
Figure 14 - Continuous Transient Operation Current Waveform
Pulse
In this mode, the electronic load generates a transient pulse of programmable width when
pulse transient operation is in effect.
In pulse mode, you can set A/B level, the pulse width, and A/B slew rate. The electronic load
will automatically switch to A level after maintaining A width time. Then it will switch to B level.
The electronic load will not switch to A level again until the instrument receives a trigger signal.
The following figure shows the current waveform in pulse transient operation.
Figure 15 - Pulse Transient Operation
Toggle
In toggle mode, the electronic load will switch between A level and B level when receiving a
trigger signal. The following picture shows the current waveform in toggle transient operation.