
39
Chapter 5
Calibration
5.1 Introduction
Calibration is an adjustment of your sound level meter to measure and display correct values.
The sensitivity of the microphone as well as the response of the electronic circuitry can vary
slightly over time, or could be affected by environmental conditions such as temperature and
humidity. While you are unlikely to ever experience a large drift or change in sensitivity with
the analyzer, it is nevertheless good practice to perform regular calibrations, normally before
and after each set of measurements. Often calibration is required by measurement standards,
such as IEC 61672–1.
If you want your analyzer to remind you when the next Accredited/Traceable calibration is due,
you can set it up on the Reminder tab of the Calibration screen. See section 5.6.
You can check the complete measurement chain during long-term measurement using Charge
Injection Calibration. See section 5.7.
5.2 Calibration Types
Acoustic Calibration
Acoustic calibration is the preferred calibration method, particularly when standards and
regulations require calibration before a measurement. The method involves applying an
acoustic signal of known magnitude and frequency to the microphone, which calibrates all the
instrument’s components (microphone, preamplifier and electrical circuitry).
Electrical Calibration
If no sound level calibrator is available (or a known amplification is introduced, i.e., by
analyzing a tape recorded signal) then you can type the sensitivity directly into the Sensitivity
field. The analyzer will be regarded as uncalibrated, and the text ‘Uncal.’ will appear in the
status field.
Commentaires sur ces manuels