98
Hand-held Analyzer 2250 Light – User Manual
Check the About menu to see whether you have the license to run the Logging module. (The
About Menu is accessed from built-in help – tap on the shortcut bar, then select About.) See
Chapter 9 for instructions on installing the license for the Logging Module.
11.1 Setting up the Instrument
1) Select the LOGGING 1/1-OCTAVE, LOGGING 1/3-OCTAVE or LOGGING SLM
Project Template. (section 3.3.1 for more details on templates.) The Project Template is
displayed on the black banner at the top of the screen. If this banner does not display
LOGGING 1/1-OCTAVE, LOGGING 1/3-OCTAVE or LOGGING SLM, tap the
banner and select one of the above from the drop-down that appears.
Note: The Logging Project Template assumes you have a license for the Frequency
Analysis Software. If not, then select the LOGGING SLM Project template instead.
2) Depending on your memory device:
• Insert an SD memory card in a SD card slot at the bottom of the analyzer
• If you have hardware versions 1 – 3, you can insert a CF memory card in the CF card
slot at the bottom of the analyzer
• With hardware version 4, you can insert a USB Type A stick at the bottom of the
analyzer
You will be notified that a memory card has been inserted – select Yes to change the
default measurement path to the memory card.
3) Tap the Main Menu icon and select Explorer from the list of options. Create a job
folder for the measurements and set the default measurement job/path as described in
Chapter 6.
Note: You cannot log data on the internal disk.
4) Tap the Main Menu icon and select Setup from the list of options. The Setup screen
will appear, see Fig.11.1. Set the Input, Frequency Settings and Statistics parameters as
required for the Sound Level Meter and Frequency Analyzer
‡
measurement, see Chapter 3
and Chapter 10 respectively. These settings are common to both the logging and the total
measurement.
‡. Requires 1/1- or 1/3-octave Frequency Analysis Software for 2250 Light BZ-7131/32