18
Hand-held Analyzer 2250 Light – User Manual
Fig.3.2
Changing the
project template
2) Tap on Measurement Mode in the view area and set to Manual (Fig.3.3).
Note: The bar graph showing the sound pressure level L
AF
is now live. This is because
L
AF
is an instantaneous value, always available for display. Other parameters, such as
L
Aeq
is a measured value that needs to be averaged over a period of time and cannot be
displayed before you have started a measurement using the Start/Pause pushbutton .
Fig.3.3
Setting the
measurement mode
3) Press the Start/Pause pushbutton to start the measurement.
Note: Start briefly appears on the screen as feedback when you press the Start/Pause
pushbutton.
Notice the Running icon on the screen and monitor the red, yellow, green status
indicators around the Start/Pause pushbutton while you are measuring.