Hand-held Analyzer Types 2250 and 2270 – User Manual
D.10 Notification Settings
Table D.10 Notification settings
Parameter Values Comment
Notification Disabled
E-mail
SMS
Set to Disabled to disable notifications.
Set to E-mail to send notifications as e-mail when an
alarm occurs.
Set to SMS to send notifications as SMS when an
alarm occurs.
See section 8.4 for the list of alarm types
Daily Status Enable
Disable
Set to Enable to send a notification once per day
Send at 00:00:00 to 23:59:59 Determines the time of day to send the daily status
notification
Alarm Settings
Internal Battery High
Low
Critical
A notification is sent, if internal battery enters the set
state
Trigger Input Voltage –20.0 V to +20.0 V A notification is sent, if voltage level at trigger input
drops to or below this setting.
Note: This alarm requires Setup > Input > Trigger
Input is set to Voltage for Monitoring
Disk Space 0% to 100% A notification is sent, if free disk space drops to or
below this setting
Level Trigger Events Disabled
Enabled
A notification is sent, if Level Trigger (specified in
Setup > Level Trigger in Logging Software or in
Setup > Triggers in Enhanced Logging Software) is
exceeded
Report Level
Notifications
Disable
Enable
Set to Enable to send a notification when the
measured L
Aeq
for the latest Report Period exceeds
Report Threshold Level (the functionality is only
available when running the Enhanced Logging
template)
Report Threshold
Level
–100.0 dB to 200.0 dB A notification is sent when the measured L
Aeq
(for
Microphone input) or Linear (for Acceleration and
Direct input) for the latest Report Period exceeds the
set Report Threshold Level.
Note 1: If you specify a very low Report Threshold
Level you will receive a notification at the end of each
Report Period
Note 2: For 2-channel Enhanced Logging, the Report
Threshold Level is for Ch.1