APPENDIX A
Setup Parameters
239
Stop Duration 0 to 420 s Stop trigger conditions are fulfilled when level
crosses Stop Level (in accordance with Start
Slope) for at least Stop Duration seconds – set
Stop Duration here. The Stop Duration will be
limited by the internal recording buffer size
and the Recording Quality and Resolution
(Table A.18)
Trigger Parameter Microphone:
L
Xeq
L
Yeq
L
Aeq,T1,mov
L
Aeq,T2,mov
L
Ceq,T1,mov
L
Ceq,T2,mov
L
Geq
L
G10max
L
G10min
L
Vpeak
L
XFmax
L
XSmax
L
XImax
L
YFmax
L
YSmax
L
YImax
L
XFmin
L
XSmin
L
XImin
L
YFmin
L
YSmin
L
YImin
L
XF(SPL)
L
XS(SPL)
L
XI(SPL)
L
YF(SPL)
L
YS(SPL)
L
YI(SPL
L
Weq
(f1-f2)
(Values continue on
next page)
Select which parameter to monitor for the level
trigger. The parameter is based on 1 second
measurements and checked every second
regardless of the logging period and logged
parameters
X = frequency weightings A or B (controlled by
Setup > Frequency Settings > BB (ex. peak)
parameter)
Y = frequency weightings C or Z (controlled by
Setup > Frequency Settings > BB (ex. peak)
parameter)
V = frequency weightings A, B, C or Z
(controlled by Setup > Frequency
Settings > BB Peak parameter)
W = frequency weightings A, B, C or Z
(controlled bySetup > Frequency
Settings > Spectrum parameter)
(Comments continue on next page)
Table A.11 (Cont.) Trigger parameters
*
Parameter Values Comment